Each set of FACA will come with one video to train in the installation, operation, and maintenance of the biochemistry analyzer.
System Functions
Type: Fully Automated, Random Access Biochemistry Analyzer.
Throughput: 200 tests, up to 320 tests/hr with ISE (optional).
Test Mode: End point, Kinetics, Fixed Time, lmmunoturbidimetry, Monochromatic, Bi-chromatic and Multi-Standard Analysis, Linearity and Nonlinearity Calibration, Single and Dual Reagents can be used.
Test Setting: Programmable for single-parameter, multi-parameter, combined parameter, batch test.
Parameter Setting: Completely open for editing test 3ertyuio
1 parameters and calculation parameters.
Sample System
Sample Tray: 44 sample positions for primary tubes, test tubes and various sample cups, Whole sample tray is removable and replaceable. Any position can be aligned as the emergency test position.
Sample Volume: 3 µL – 30 µL, programmed by 0.1 µL step.
Sample Handling: Dilute and retest automatically when the sample' s concentration is beyond limitation.
Sample Probe: Liquid level detection and collision protection.
Probe cleaning: Automatic washing for both inside and outside, carry over <0.1%.
Syringe: High performance syringe, durable and maintenance free.
Optional item: Barcode reader.
Reagent System
Reagent Tray: 44 positions for Single and Dual reagents, reagent positions can be aligned freely.
Refrigeration: Refrigerated reagent compartment, 24 hours non-stop cooling with Peltier element to maintain temperature 4-15 ℃.
Reagent Bottle Volume: 45ml and 15ml, Slope bottom design.
Reagent volume: 150 µL - 300 µL, prograrnmed by 0.1µ1 step.
Reagent Probe: Automatically detecting the liquid level; collision protection.
Reagent pre-heating function, inventory checking.
Real time monitoring of reagent residual volume and auto alarm.
Probe Cleaning: Washing for both inside and outside, carry over <0. 1%.
Reaction System
Reaction Tray: 44 reaction cuvettes with 6mm optical path, good UV penetrability, low consumable, low cost.
Reaction Technology: Back – Dividing – Light – Technology.
Reaction Volume: 250 µL - 300 µL.
Temperature: 37℃, with +0.1℃ fluctuation.
Mixing System: Independent mixer.
Cleaning: Supports automatic cleaning of cuvettes and allows for complete drainage to reduce carry over.
Water consumption: 3L/hour.
Liquid Detection: Detergent, distilled water, waste liquid automatic detection and alarm.
Optical System
Light Source: Halogen lamp 6V/ 10W, life span > 3000 hours.
Wavelength: 340nm, 405nm, 450nm, 492nm, 505ml, 546nm, 578nm, 620nm, 670nm, 700nm
Wavelength Accuracy: +2nm.
Detector: Photoelectron Diode Array.
Absorbance Range: 0.000 Abs -3.000 Abs.
Calibration and Quality Control
Calibration Mode: Linear (one-point, two-point and rnulti-point ), Logit-Log4P, Logit-Log5P, Spline, Exponential, Polynomial, Parabola.
QC Rules: Westgard Multi-rule.
QC Mode: Intra-day and inter-day QC.
Operating System
Computer: Windows 2000 System or above.
Data Management: Large capacity for results and reaction curves to be displayed, saved and printed.
Report: Edit and print complete reports, format editable, supports network printing.
Port: RS-232.
Filters:
FACA200 will have 10 filters 340,405,450,492,505,546,578,620,670,700nm
Working Conditions
Power: 220V ±22V, 50/60Hz ± 1Hz
Power Consumption: <400VA
Environmental Temperature: 15℃ - 30℃
Humidity: <85%
Dimension: 85x75x75 cm (LxWxH)
Weight: 145 Kgs
QC control function, based on Levy-Jennings chart analysis
This instrument is open system and can use any common brand reagents. |