This website is better viewed at 1024 x 768 resolution. |
|
|
Fully Automatic 3-Part 23 Parameters Hematology Analyzer
H-7022 Touch Screen
(Less Service, More Stable)
PDF Brochure - Page 1
PDF Brochure - Page 2
PDF Brochure - Page 3
H-7022 Touch Screen 3-Part 23 Parameters Hematology Analyzer (Less Service, More Stable) is a 3-part differential of WBC blood cell counter. Its advantages include digital technology, auto-relocation separatrix, intelligent flush clog, dual measuring technology (with time and volume) to ensure precise test results.
Automated and User-Friendly Interface
- 8-inch LCD display, resolution: 800x600
- Real-time alert for abnormal results
- USB keyboard and mouse operation
- Based on Linux
- Easy to search and edit patient information
- Storage capacity: 50,000 results including histograms
- Optional printout for formats
- Real-time guidelines for easy operation
Accurate Results and Low Sample Consumption
- Accurate analysis technology with low sample consumption
- Venous mode: 9.6ul
- Capillary mode: 9.6ul
- Pre-diluted mode: 20ul
Easy Maintenance and Environmentally-Friendly System
- Adjustable sample probe for various cuvettes
- With liquid sensor to alarm low reagent
- Cyanide-free Lyse
- Automated and Systematic diagnosis and maintenance with alerts
- Liquid tubing and electronic circuit separately for easy maintenance
- Closed reagent tubing to avoid contamination
Perfect QC System
- Calibration mode: manual and auto
- QC mode: L-J, SD, CV, QC histogram
Advanced Technology
- Unique sweep flow technology to avoid recounting
- Auto-relocation separatrix to exact the differential of WBC
- Intelligent PLT histogram fit technology to improve the precision of PLT
- Auto sample probe cleaning
- Auto standby and wakeup
- High voltage burn, soaking and flush clog with real time monitor
Technical Specifications
Operation Mode: |
USB keyboard & mouse. Based on Linux system |
Parameters: |
3-part differential of WBC; 20 parameters + 3 color histograms (WBC/RBC/PLT) |
Sample Consumption: |
Venous Mode: 9.6µL, Capillary Mode: 9.6µL, Pre-diluted Mode: 20µL |
Working Mode: |
Double passages counting detection + independent HGB measurement system; 24 hours with auto standby |
Language: |
Multi-language supported |
Throughput: |
≥60 samples/hour |
Storage: |
50,000 sample results with histograms, record supported export and backup L-J,SD,CV,QC histogram
|
QC mode: |
L-J, SD, CV, QC histogram |
Cross Contamination Ratio: |
WBC/RBC/HGB<0. 5%, PLT<1% |
Performance: |
Parameter Repeatability (CV%) |
Linear Range |
WBC≤2.5% |
0.00-99.9X109/L |
RBC≤2.0% |
0.00-99.9X1012/L |
HGB≤1.5% |
0-300g/L |
MCV≤0.4% |
40-150fL |
PLT≤5.0% |
0.00-99.9X109/L |
|
Display: |
8" LCD, resolution :800 x 600 |
Reference Value: |
Baby, Kid, Woman, Man, Normal and Customization |
Interface: |
4 xUSB, 1 x ethernet, 1 x VGA (for external monitor) |
Printer: |
Bulit-in thermal printer, supports external laser and inkjet printer, 5 kinds of report formats supported |
Environment Requirement: |
Temperature: 15°C-35°C; Humidity: 10%-90% |
Power Supply: |
AC100-240V 50/60Hz |
Top
of the Page
Labomed, Inc. is certified by ISO 9001-2013, has CE Conformity
and is FDA Licensed.
click here to fill out the Information Request Form
Labomed, Inc. Mailing Address: 2728 S La Cienega Blvd.
Los Angeles, CA 90034 U.S.A TEL (310) 202-0814 FAX (310) 202-7286
Send mail to Company Webmaster with
questions or comments about this web site.
Copyright © 2008
|